Evaluation of three handheld lactate meters for measuring hemolymph l-lactate in decapod crustaceans

The aim of this study was to determine the reliability and accuracy of 3 commercially available handheld lactate meters when measuring hemolymph L-lactate in decapod crustaceans. The instruments tested were Accutrend Plus (AP), Lactate Scout (LS), and Lactate Plus (LP), and all models were tested in triplicate and compared with a commonly used enzymatic kit for L-lactate quantification (TB-kit). The Norway lobster Nephrops norwegicus and the European lobster Hommarus gammarus were used as model organisms, with emersion as an invoker of hemolymph L-lactate. Between-method comparisons were investigated by determining correlations (Pearson's r) and concordance (concordance correlation coefficient (CCC) and Bland-Altmand's 95% limit of agreement (95% LOA)) between handheld instruments and TB-kit. Within-instrument variation was tested by calculating the coefficient of variation (CV), and the average absolute deviation from the mean (AD).

The within-instrument variation was low for all models (CV, 3.8–6.1%; AD, <0.3 mM), showing high reproducibility of technical replicates. A significant correlation with TB-kit was found for all handheld instruments (AP, r = 0.970; LS, r =; 0.956; LP, r = 0.950). For both AP and LS, there was a moderate and significant concordance with TB-kit (AP, CCC = 0.945; LS, CCC = 0.940). The mean difference from TB-kit was -1.42 mM (95% LOA,-4.50–1.66 mM) for AP and 1.19 mM (95% LOA,-3.13–5.51 mM) for LP. Emersion experiments with the European lobster showed that AP and LS were in line with TB-kit in terms of detecting differences in L-lactate levels between groups of animals. As a result of the high frequency of error readings from LP, this instrument was not considered reliable when measuring L-lactate in decapod crustaceans.

In conclusion the handheld instruments AP and LS are found to be reliable instruments when measuring hemolymph L-lactate in decapod crustaceans. Comparison of results from different methods should be avoided, and in cases when accurate absolute measurement of L-lactate is needed, methods allowing for greater accuracy and resolution should be used.

Publisert:
2014
Forfatter(e):
No items found.
Snorre Bakke,
Utgiver:
Møreforskning
Journal of Shellfish Research
Oppdragsgiver:
Rapportnummer:
Kategori:
Vitenskapelig artikkel

Relaterte prosjekter

No items found.
Har du spørsmål om vår forskning eller ønsker et samarbeid?

Ønsker du å samarbeide med Møreforsking, eller har du spørsmål om våre publikasjoner eller prosjekter? Ta gjerne kontakt, så hjelper vi deg videre og finner den løsningen som passer best. Våre forskere har bred kompetanse og lang erfaring innen ulike marine fagområder.